IATT - Application Guide


GAS CHROMATOGRAPHY
Instrument Gas used Purity Flow rate Product

 

 

GC - FID

Hydrogen for detector Hydrogen high purity 30 to 50 mL/min Hydrogen
Zero air for detector Without hydrocarbon 300 to 500 mL/min Air Zero
Hydrogen for Carrier gas Hydrogen high purity 10 mL/min Hydrogen
Nitrogen for Carrier gas Nitrogen high purity, Zero N2 20 to 50 mL/min Nitrogen
Nitrogen for Make-up Nitrogen high purity, Zero N2 30 to 50 mL/min Nitrogen
 

 

 

GC - FPD

Hydrogen for detector Hydrogen high purity 60 to 90 mL/min Hydrogen
Zero air for detector Without hydrocarbon 90 to 120 mL/min Zero Air
Hydrogen for Carrier gas Hydrogen high purity 10 mL/min Hydrogen
Nitrogen for Carrier gas Nitrogen high purity, Zero N2 20 to 50 mL/min Nitrogen
Nitrogen for Make-up Nitrogen high purity, Zero N2 30 to 50 mL/min Nitrogen
 

 

GC - NPD

Hydrogen for detector Hydrogen high purity 50 mL/min Hydrogen
Nitrogen for Carrier gas Nitrogen high purity, Zero N2 20 to 50 mL/min Nitrogen
Nitrogen for Make-up Nitrogen high purity, Zero N2 30 to 50 mL/min Nitrogen
 

GC - ECD

Nitrogen for Carrier gas Nitrogen high purity, Zero N2 30 mL/min Nitrogen
Nitrogen for Make-up Nitrogen high purity, Zero N2 60 mL/min Nitrogen
 
GC - TCD Hydrogen for Carrier gas Hydrogen high purity 100 mL/min Hydrogen
Nitrogen for Carrier gas Nitrogen high purity 100 mL/min Nitrogen
 
GC - FPC Nitrogen for Make-up Nitrogen high purity, Zero N2 30 to 50 mL/min Nitrogen
 
GC - ATD Zero air -0,05ppm hydrocarbons < 2 L/min Zero Air
 
GC - AED Nitrogen for Carrier gas Nitrogen high purity, Zero N2 < 1 L/min Nitrogen
 
GC - ELCD Hydrogen for reaction gas Hydrogen high purity 70 to 200 mL/min Hydrogen
 
GC - IR Air without CO2 for FT-IR Dry air without CO2 14 to 85L/min Dryer without CO2
 
GC - MS Carrier gas Nitrogen or hydrogen high purity - Zero Nitrogen
or Hydrogen
 
FAST - GC Hydrogen high pressure Hydrogen high pressure - Hydrogen

 

ANALYSERS
Instrument Gas used Purity Flow rate Product
COT Dry air or nitrogen for carrier gas and combustion gas CO2-free dry air, without hydrocarbon 100 to 700 mL/min Air TOC series
High purity nitrogen 50 to 700 mL/min Nitrogen
 
CO2 Analyser Calibration CO2-free dry air, without hydrocarbon 550 to 1000 mL/min UZAG or Air TOC series
 
DSC Dry air Dry air 100 mL/min Air Dryer
Curtain gas High purity nitrogen 100 mL/min Nitrogen
 
TGA Dry air or Nitrogen for furnace Nitrogen : 99.5%-3b or dry Air 4-200 mL/min Air Dryer, Nitrogen
 
DMA Dry air or Nitrogen for furnace and drive bearings High purity nitrogen or dry air - Air Dryer, Nitrogen
 
THA Hydrogen and Zero air for FID detector Hydrogen and air without hydrocarbon - Zero Air
and Hydrogen
 
TMA Nitrogen for furnace High purity nitrogen <1L/min Nitrogen
 
TOD - High purity nitrogen <600ml/min Nitrogen

 

LMCS
Instrument Gas used Purity Flow rate Product
LCMS / MS, TOF Nitrogen for
curtain gas
99% Up to 64 L/min (Depending
on LCMS models)
Nitrogen
 
AB SCIEX Curtain Gas Nitrogen only Curtain Gas :
12 L/min @ 5.52 bar
Nitrogen
Source Gas Nitrogen only Source Gas :
24 L/min @ 7.58 bar
Nitrogen
Exhaust Gas Nitrogen only Exhaust Gas :
8 L/min @ 4.14 bar
Nitrogen
Curtain Gas Nitrogen and dry air Curtain Gas (Nitrogen)
12 L/min @ 5.52 bar
-
Source Gas Nitrogen and dry air Source Gas (Dry Air)
24 L/min @ 7.58 bar
-
Exhaust Gas Nitrogen and dry air Exhaust Gas (Dry Air)
8 L/min @ 4.14 bar
-

 

SPECTROSCOPY
Instrument Gas used Purity Flow rate Product
FT-IR Purge gas for sample compartment, optics, air bearing and microscope Dry Air without CO2 15 to 85 L/min CO2 Removal
Air Dryer
RMN Air for lifting spinning and ejecting Dry Air 300 L/min Dry Air
ICP Nitrogen or Zero Nitrogen for purge gas > 99,9995 % 1 to 9 L/min Nitrogen
ICP-MS Collision gas in plasma chamber > 99,99999 % 80 ml/min Hydrogen
AA Air for oxidant gas Dry Air 28 to 200 L/min Air Dryer

 

DETECTORS
Instrument Gas used Purity Flow rate Product
CORONA Nitrogen 98% Up to 6 L/min Nitrogen
 
ELSD Air
for nebulisation
Dry air 2-6 L/min Dryer
Nitrogen
for nebulisation
Nitrogen 4 –8 L/min Nitrogen

 

OTHER APPLICATIONS
Instrument Gas used Purity Flow rate
Solvent
evaporation
Nitrogen for solvent evaporation from 95 % to 99 % Approx. 130 L/min Nitrogen
 
TurboVaps Air for pneumatic control 28 L/min Dry Air
Nitrogen for auto-sampler 550 mL/min Nitrogen
 
Circular dichroism Nitrogen @ 99.999 % up 15 L/min Nitrogen
Laser Diffraction Clean and dry air for nebulizing gas Depend on model Air Dryer
Hydrogenation Hydrogen Depend on model Hydrogen
Glove box / cover Nitrogen or dry air Depend on model Nitrogen or Dry Air
Fuel cells Hydrogen High purity Depend on model Hydrogen
Rheometer Dry Air Depend on model Dry Air
ASE DIONEX Nitrogen @ 10 bars 20 L/min Nitrogen